Pure water optical dissolved oxygen sensor designed specifically for demanding power plant and microelectronics applications
Fast Response, Reduced Maintenance
METTLER TOLEDO Thornton's Pure Water Optical Dissolved Oxygen Sensor with ISM has been designed specifically for demanding power plant and microelectronics applications.
Low maintenance optical measurement technology provides exceptional measurement accuracy, stability, and speed of response.
Paired with the M800 multi-parameter transmitter, this system offers unequalled performance and reliability.
Applications
Power plant cycle chemistry monitoringOperating range | 0 - 5000ppb |
Accuracy | 2 % + 2 ppb |
Response time | 98% of final value in < 20 s |
Measuring temperature range | 10 – 50 °C (50 – 122 °F) |
Environmental temperature range | 0 – 121 °C (32 – 250 °F) |
Operating pressure | 0.2 – 12 bar (2.9 – 174 psi absolute) |
Mechanical pressure resistance | Maximum 12 bar (174 psi absolute) |
Segment /Application | The most demanding low ppb power and microelectronics pure water applications |
Detection limit | .8ppb |