New Product

106-H
Industrial Ozone MonitorTM

Features

  • Recommended for Industrial Ozone Applications for Concentrations in the Range 0-1,000 ppm
  • Measurement based on UV absorption
  • Low power consumption; can be battery operated
  • Internal data logger with real-time clock

Get a quote

The Model 106 series of ozone monitors was designed specifically for the industrial ozone industry to cover four different ozone concentration ranges by varying the optical path length. The ranges are: -L (low, 0-100 ppm); -M (medium, 0-1,000 ppm); -MH (medium high, 0-10,000 ppm, 0-1 vol%) and -H (0-20 wt%, 0-14 vol%). The Model 106 series is designed as an “ozone monitor on a board” in which nearly all of the components are mounted directly to the printed circuit board with very few wire connections, making these instruments highly robust and very easy to service. As seen in the images above, all models are available in a standard enclosure, an Industrial enclosure, or with no enclosure for OEM applications for those who want to mount the instrument in their own enclosure or use it as a component of a larger system. For more details on OEM applications see: Model OEM-106. Configurations for multi-channel sampling are available for Models 106-L, 106-M, and 106-MH.

The Model 106-M Ozone Monitor is designed for measurements of ozone in the range 0-1,000 ppm with a resolution of 0.01 ppm. A common application is off-gas analysis in water treatment plants, before and after ozone destruction.

Application Note

The 0-1000 ppm range of the Model 106-M makes it useful for many industrial and research applications. Below is an example of its use in characterizing lamps used in oxidation flow reactors.

Download Study: The Model 106-M Used in an Oxidation Flow Reactor Study

Features:

  • Measurement based on UV absorption
  • Low power consumption; can be battery operated (optional external lithium-ion battery)
  • Internal data logger with real-time clock
  • 2-s measurement interval
  • On-board microprocessor with interactive menus includes data averaging options of 10 s, 1 m, 5 m, 1 hr, and user-defined adaptive filter
  • USB and RS-232 output of time/date, O3 concentration, internal temperature and pressure
  • Analog output (0-2.5 V and 4-20 mA) of ozone concentration in user-selected units and scaling factors
  • Two 2-level relays for control purposes (e.g., control of ozone source or turn warning light on and off); Four relays on Industrial models
  • Long-life pump (15,000 hr)
  • Bluetooth option for wireless data transmission
  • Optional configurations for 3-channel or 6-channel air sampling for Model 106-L, -M, and -MH

Industrial Technical Services from Mississauga, Ontario is a single source Instrumentation and Automation company. We provide certified calibrations, PLC programming, custom controls and Panels.

Specifications & Spare Parts

Measurement Principle UV Absorption at 254 nm, single beam
Measurement Interval 2 s
Linear Dynamic Range 0-1,000 ppm
Resolution 0.01 ppm
Precision (1σ for 10-s average; aka rms noise) Greater of 0.01 ppm or 2% of reading
Limit of Detection (10-s average, 2σ) 0.02 ppm
Accuracy Greater of 0.01 ppm or 2% of reading
Baseline Drift < 0.01 ppm/day, < 0.03 ppm/year
Sensitivity Drift < 1%/day, < 3%/year
Calibration NIST Traceable; Annual calibration recommended
Measurement Time and Frequency 2 s, 0.5 Hz
Data Averaging Options 10 s, 1 min, 5 min, 1 hr
Response Time, 100% of Step Change For 2-s output: 4 s, 2 data points
For 10-s output: 20 s, 2 data points
Adaptive Filter Available; user-defined parameters
Data Logger Capacity 32,736 lines (10 s avg. = 3.8 days; 5 min avg = 113 days)
Data Transfer Baud Rates 2400, 4800, 19200
Ozone Units ppb, pphm, ppm, µg m-3, mg m-3
Temperature Units oC, K
Pressure Units mbar, torr
T and P Corrected Yes
DewLine™ for Humidity Control Yes
Operating Temperature Range 0 to 50oC
Operating Altitude Range 0 to 5 km
Power Requirements 100-240 VAC, 50/60 Hz
11-28 V DC, nominally 500 mA at 12 V DC, 6 watt
Flow Rate Minimum required: 0.3 L/min; Nominal: 1 L/min; Maximum: 1.5 L/min
Digital Data Outputs USB, RS232, LCD display
Analog Data Outputs 0-2.5 V Analog, 4-20 mA, user-scalable in menu
Relays with 2 Setpoints Two available:
Relay 1 responds based on user’s ozone set points
Relay 2 reponds based on user’s ozone set points OR responds based on diagnostics (T, P, flow, lamp voltage)
Four relays on Industrial models
Bluetooth Option Available Yes
Flow-Through Option Available Yes
Multi-Channel Options Available Yes, 3-channel and 6-channel configurations on Model 106-L, -M, and -MH
Size Standard: 3.6 x 7.9 x 9.4 inches (9 x 20 x 24 cm)
OEM: 2.5 x 7 x 9 in (6.4 x 17.8 x 22.9 cm)
Industrial: 16 x 14 x 9.3 in (40.7 x 35.7 x 23.6 cm) H x W x D
Weight Standard: 3.9 lb (1.8 kg)
OEM: 2.5 lb (1.1 kg)
Industrial: ~14 lb (6.4 kg) depending on multi-channel options
Options Battery, Particle Filter, Bluetooth, Exhaust Port, Flow-Through Configuration, Multi-Channel Configurations

Product Tags

, , , , , , ,