New Product

BetaGauge 311A-EX/321A-EX

Intrinsically Safe Pressure Calibrator

Get a quote

Intrinsically Safe Pressure Calibrator

Whichever you choose, the single sensor BetaGauge 311A-EX, or the dual sensor BetaGauge 321A-EX, you’ll have everything you need for calibrating pressure anywhere. Gas custody transfer is the ideal application for the accuracy and capabilities of the BetaGau­­ge 321A-EX. Create a custom configuration by selecting from any two ranges. The BetaGauge 311A-EX can be configured by selecting from any range. Both are supplied with soft case, AA batteries, manual, NIST-traceable certificate, and test leads. Both are available in kits including the Pressure Calibrator, Pt100 RTD Probe, AA batteries, manual, NIST-traceable certificate, test leads, and hard-sided carrying case.

Approval Details
Ex ia IIB T3 Gb (Ta=–10… +45°C)
KEMA 10 ATEX 0168X 0344
Ex ia IIB T3 Gb (Ta=–10… +45°C)
II 2 G IECEx CSA 10.0013X

General Features

  • Single or dual pressure sensors with up to ±0.025% F.S. accuracy
  • 27 available pressure ranges
  • Temperature-compensation ensures accuracy in the field applications
  • Pt100 RTD input for temperature measurement, accurate to 0.1°C (0.2°F)
  • Measure 4-20 mA input
  • ClearBrite™ graphic style LCD with backlight can be configured to display up to 3 inputs simultaneously (i.e. pressure 1, pressure 2, and RTD or any combination of inputs)
  • Ideal for gas flow calibration (custody transfer) applications
  • Pressure displayed in the user’s choice of 19 engineering units
  • Switch test on Pressure 1 or Pressure 2
  • Operating mode layers support simple measurements to complex applications
  • Up to five frequently used setups can be stored; last setup automatically recalled on power-up
  • % Error and damping functions
  • Small, rugged compact design operates on four (4) Alkaline AA batteries

Specifications & Spare Parts

Specifications (15 °C to 35 °C, unless otherwise noted)
Ranges
Available Pressure
Isolated Gauge (PSIG): 0 to 1000 (0 to 70 Bar), 0 to 1500 (0 to 100 Bar),
0 to 3000 (0 to 200 Bar), 0 to 5000 (0 to 340 Bar),
0 to 10000 (0 to 700 Bar)
Non Isolated Compound (PSIG): -0.4 to 0.4 (-20 to 20 mBar), -1 to 1 (-70 to 70 mBar),
-5 to 5 (-350 to 350 mBar), -7.2 to 7.2 (-500 to 500 mBar),
-15 to 15 (-1 to 1 Bar), -15 to 30 (-1 to 2 Bar)
Isolated Compound (PSIG): -12 to 100 (-0.8 to 7 Bar),
-12 to 300 (-0.8 to 20 Bar),
-12 to 500 (-0.8 to 35 Bar)
Isolated Absolute (PSIA): 0 to 15 (0 to 1 Bar), 0 to 30 (0 to 2 Bar), 0 to 100 (0 to 7 Bar),
0 to 300 (0 to 20 Bar)
Non Isolated Differential (PSID): -0.4 to 0.4 (-20 to 20 mBar), -1 to 1 (-70 to 70 mBar),
-5 to 5 (-350 to 350 mBar), -12 to 30 (-0.8 to 2 Bar),
-12 to 50 (-0.8 to 3.5 Bar), -12 to 100 (-0.8 to 7 Bar)
mA 0 to 24.000 mA
RTD -40°C to 105°C (-40°F to 220°F)
Engineering Units
psi, bar, mbar, kPa, kgcm2, cmH2O@4°C, cmH2O@20°C,
mmH2O@4°C, mmH2O@20°C, inH2O@4°C, inH2O@20°C,
inH2O@60°F, mmHg@0°C, inHg@0°C, ftH2O@4°C,
ftH2O@20°C, ftH2O@60°F
Plus up to 2 user specified custom units
Accuracy
Pressure
0.4 psi ±0.1 % FS
1.0 psi ±0.05 % FS
15 to 3000 psi ±0.025 % FS
5, 7.2, 5000, 10000 psi ±0.035 % FS
mA ±0.015% of rdg ±0.002mA
RTD input range (Pt100-385) ±.01°C
Temperature Effect
No effect on accuracy on all functions from 15°C to 35°C
Add ±0.002% F.S./°C for temps outside of 15°C to 35°C
General
Instrument Setup Recall 5; last used on power-up
Environmental
Operating Temperature -10 °C to +50 °C
Storage Temperature -20 °C to +60 °C
Power Requirements 6.0 VDC
Battery Four (4) standard AA cells
Battery Life >35 hours, typical usage
Physical
Dimensions 8.3” H x 3.9”W x 1.8”D (21.082 x 9.906 x 4.572 cm)
Weight 1 lb, 4 oz. (0.567 kg)
Connectors/Ports Pressure
Two, 1/8” NPT
RTD