New Product
Ozone (Ambient) / Chemical & Petrochemical / Pharma / Water & Waste Water
Model 205 Dual Beam Ozone Monitor High
Ozone (Industrial) / Chemical & Petrochemical / Pharma / Water & Waste Water
106-MH Ozone MonitorTM (Industrial)
Ozone (Industrial) / Automotive & Manufacturing / Chemical & Petrochemical / Dairy / Food & Beverage / Lab & Gas / Pharma / Plastics / Power & Steam
106-HIndustrial Ozone MonitorTM
InPro6970i Trace Optical Oxygen Sensor
InPro 6970i trace optical oxygen sensor designed for brewery applications
Related Products
Optical oxygen sensor for breweries.
The new InPro6970i designed for brewery application sets a new standard for operational availability and measurement quality of your oxygen loop especially in the filler line. The optical oxygen sensors are easy and fast to maintain, leading to enhanced process safety and low cost of ownership. Reduced signal drift and faster response time compared to amperometric sensors guarantees maximum accuracy of the oxygen measurement and allow for enhanced process control.
Features and Benefits
| Easy operation | The InPro6970i® with its plug and measure functionality combined with fastest Opto-Cap replacement sets a new standard in ease of use. |
| High process safety | FDA compliant materials of construction and easy-to-clean high-polished surface (N5 grade) to satisfy increasing regulatory requirements. |
| Maintenance redefined with ISM | The built-in ISM (Intelligent Sensor Management) provides the user important information about the sensor wear and enables predictive maintenance. |
| Enhanced process control | Highest measurement performance in filler lines. |
Description
The light source and the detector are both integrated in the sensor head. The unique ISM (Intelligent Sensor Management) technology is implemented in all optical sensors from METTLER TOLEDO. All system data including calibration and diagnostic information are stored in the sensor.
The METTLER TOLEDO sensors InPro6970i are based on the so call fluorescence quenching. The signal of the fluorescence is compared with a reference signal and the oxygen concentration is calculated from the time shift between these two signals. The oxygen concentration is displayed on the transmitter.
Specifications & Spare Parts
| Segment /Application | Brewery applications with fastest response time requirement |
| Accuracy | 1% + 2 ppb |
| Material in contact with medium | stainless steel (AISI316L) with 3.1B certificates |
| Mechanical pressure resistance | max. 12 bar (174 psi) absolute |
| Detection limit | 2 ppb to 2000 ppb |
| Connector | 5 Pin |
| Sterilizable | yes |
| ISM | yes |
| Insertion length | for 12 mm: 120, 220, 320 mm |