PD67000 Vantageview NEMA 4X Loop-Powered Process & Level Meter
The Vantageview PD6700 is one of Precision Digital’s most advanced and feature-packed plastic, field-mounted, loop-powered indicators. The fact that this meter is loop-powered means that there is no need to run additional, costly power lines. The PD6700 is built in a NEMA 4X and IP65 rated plastic enclosure which can stand up to even the harshest environmental conditions. It features a dual-line display that can be easily scaled in the field for any 4-20 mA analog input. The display has a five digit upper display, a seven alpha-numeric character lower display, and comes with a backlight standard. Two features that really make the PD6700 stand out are its wide viewing angle display and SafeTouch® through-window buttons. These buttons allow the meter to be programmed and operated through the cover window, thus eliminating the need to remove the cover in dirty or wet environments.
Features
|
Specifications & Spare Parts
General Specifications
| Display | Upper: Five digits (-9999 to 99999) 0.70″ (17.8 mm) high, 7-segment, automatic lead zero blanking; -0L1: 0.6″ high with level bar graph. Lower: Seven characters 0.4″ (10.2 mm) high, 14 segment alphanumeric. Symbols: High & Low alarm, Password Lock. Backlight: White |
| Decimal Point | Upper process display has up to four decimal places or none: d.dddd, d.ddd, d.dd, d.d, or ddddd |
| Display Update Rate | Ambient > -25°C: 2 Updates/Second. Ambient < -25°C: 1 Update/5 Seconds |
| Display Orientation | Display may be mounted at 90° increments up to 270° from default orientation. |
| Overrange | Display flashes 99999 |
| Underrange | Display flashes -9999 |
| Programming Method | Four SafeTouch® through-window buttons when cover is installed. Four internal pushbuttons when cover is removed. |
| Noise Filter | Programmable – Lo, Med, Hi, or Off |
| Recalibration | Recalibration is recommended at least every 12 months. |
| Max/Min Display | Max/Min readings reached by the process are stored until reset by the user or until power to the meter is turned off. |
| Password | Programmable password restricts modification of programmed settings. |
| Advanced Functions | Live input calibration, linearization, square root, or programmable exponent |
| Alarm Indication | Flashing display plus HI/LO indicators |
| Non-Volatile Memory | All programmed settings are stored in non-volatile memory for a minimum of ten years if power is lost. |
| Normal Mode Rejection | 64 dB at 50/60 Hz |
| Operating Temperature Range | -40 to 75°C. |
| Storage Temperature Range | -40 to 75°C. |
| Relative Humidity | 0 to 90% non-condensing |
| Connections | Screw terminals accept 12 to 22 AWG wire |
| Enclosure | NEMA 4X, IP65 plastic field enclosure. Color: grey. Material: Polycarbonate with UV Stabilizer. Three ¾” NPT threaded conduit openings. Two ¾” NPT plastic conduit plugs, with 1.29″ wrenching flats and a screwdriver slot, are included. |
| Mounting | May be mounted directly to conduit. Two slotted flanges for wall mounting or NPS 1½” to 2½” or DN 40 to 65 mm pipe mounting. |
| Overall Dimensions | 5.67″ x 5.25″ x 4.18″ (W x H x D) (144 mm x 133 mm x 106 mm) |
| Weight | 1.65 lbs (26.4 oz, 0.75 kg) |
| Warranty | 3 years parts and labor |
Input Specifications
| Input Range | 4-20 mA | ||||
| Accuracy | ±0.03% of calibrated span ±1 count, square root & programmable exponent accuracy range: 10-100% of calibrated span. | ||||
| Temperature Drift | 50 PPM/°C from -40 to 75°C ambient | ||||
| Decimal Point | User selectable decimal point | ||||
| Calibration Range | An error message will appear if input 1 and input 2 signals are too close together. Input Range: 4-20 mA. Minimum Span Input 1 & Input 2: 0.10 mA | ||||
| Maximum Voltage Drop | 3.0 VDC @ 20 mA without loop-powered backlight. 6.0 VDC @ 20 mA with loop-powered backlight | ||||
| Equivalent Resistance | 150 Ω @ 20 mA without loop-powered backlight. 300 Ω @ 20 mA with loop-powered backlight | ||||
| Input Overload | Over current protection to 2 A max. | ||||
| External Loop Power Supply |
|
Open Collector Output (-0L1 Models Only)
| Rating | Isolated open collector, 30 VDC @ 150 mA max. |
| Alarm Output | Assign to level or volume for high or low alarm trip point. |
| Deadband | 0-100% FS, user selectable |
| Acknowledge | Front panel ENTER button and external RESET terminals resets output and screen indication. |