New Product

6000i
TOC Sensor

Features

  • Fast response time available and measurement updates every second for continuous TOC analysis.
  • Ideal for all pure water applications where rapid detection of TOC changes is critical
  • The 6000TOCi uses proven UV oxidation technology and highly accurate conductivity sensors to deliver consistent and accurate determination of organic material
  • Verifiable System Performance

Get a quote

Real-time Total Organic Carbon Control. The 6000TOCi online sensor delivers true continuous measurement for the detection of organic contamination. With an extremely rapid response to TOC changes, the 6000TOCi is ideal for all pure water applications where rapid detection of TOC changes is critical.

Real-Time Transparency of TOC Levels

Achieve clear control of your water system with the fastest response time available and measurement updates every second for continuous TOC analysis.

Conclusive Data to Support Compliance

The 6000TOCi uses proven UV oxidation technology and highly accurate conductivity sensors to deliver consistent and accurate determination of organic material.

Verifiable System Performance

Using advanced sensor diagnostics, the 6000TOCi gives you the insight needed to ensure your TOC system is always measuring effectively.

Specifications & Spare Parts

Specifications – TOC Sensor 6000i

Measuring range 0.05 – 2000 ppbC (µgC/L)
Measurement Range 0.05 ppb – 2,000 ppb
Response Time 60 s
Detection limit 0.025 ppb
Pressure Range 4 psi – 200 psi (0.3 bar – 13.6 bar)
Temperature accuracy (±) ± 0.25°C
Analysis Time Continuous
Limit of Detection 0.025 ppbC
Flow rate > 8.5 mL/min
Maintenance 6 mo UV replacement
Weight 6 kg (11.0 lb)
Repeatability ± 0.05 ppbC < 5 ppbC, ±1.0% > 5 ppbC
Temperature Sensor Pt1000 RTD, Class A
Dimensions HWD 11.9” [302.75 mm] W x 9” [229.8 mm] H x 5.7” [144.7 mm] D
Online or Portable Online
Enclosure Material Powder-coated aluminum back enclosure and polystyrene resin front enclosure.
Initial Response Time < 60 seconds
Resolution 0.001 ppbC (µgC/L)
Operating Range (Temp.) 0 °C – 100 °C
Conductivity Accuracy ± 2%, 0.02 – 20 μS/cm ± 3%, 20 –100 μS/cm
Cell Constant (accuracy) ± 2%
Intelligent Sensor Management (ISM®) Yes